Note: This content is accessible to all versions of every browser. However, this browser does not seem to support current Web standards, preventing the display of our site's design details.

  

Model-based control of mechanical ventilation: design and clinical validation

Author(s):

E. Martinoni, C. Pfister, K.S. Stadler, P.M. Schumacher, D. Leibundgut, T. Bouillon, T. Böhlen, A.M. Zbinden
Conference/Journal:

British Journal of Anaesthesia, vol. 92, pp. 800-807
Abstract:

Background. We developed a model-based control system using end-tidal carbon dioxide fraction (FE'CO2) to adjust a ventilator during clinical anaesthesia. Methods. We studied 16 ASA I–II patients (mean age 38 (range 20–59) yr; weight 67 (54–87) kg) during i.v. anaesthesia for elective surgery. After periods of normal ventilation the patients were either hyper- or hypoventilated to assess precision and dynamic behaviour of the control system. These data were compared with a previous group where a fuzzy-logic controller had been used. Responses to different clinical events (invalid carbon dioxide measurement, limb tourniquet release, tube cuff leak, exhaustion of carbon dioxide absorbent, simulation of pulmonary embolism) were also noted. Results. The model-based controller correctly maintained the setpoint. No significant difference was found for the static performance between the two controllers. The dynamic response of the model-based controller was more rapid (P<0.05). The mean rise time after a setpoint increase of 1 vol% was 313 (SD 90) s and 142 (17) s for fuzzy-logic and model-based control, respectively, and after a 1 vol% decrease was 355 (127) s and 177 (36) s, respectively. The new model-based controller had a consistent response to clinical artefacts. Conclusion. A model-based FE'CO2 controller can be used in a clinical setting. It reacts appropriately to artefacts, and has a better dynamic response to setpoint changes than a previously described fuzzy-logic controller.

Year:

2004
Type of Publication:

(01)Article
Supervisor:

A.H. Glattfelder

File Download:

Request a copy of this publication.
(Uses JavaScript)
% Autogenerated BibTeX entry
@Article { MarEtal:2004:IFA_2062,
    author={E. Martinoni and C. Pfister and K.S. Stadler and P.M. Schumacher and D.
	  Leibundgut and T. Bouillon and T. B{\"o}hlen and A.M. Zbinden},
    title={{Model-based control of mechanical ventilation: design and
	  clinical validation}},
    journal={British Journal of Anaesthesia},
    year={2004},
    volume={92},
    number={},
    pages={800--807},
    month=jun,
    url={http://control.ee.ethz.ch/index.cgi?page=publications;action=details;id=2062}
}
Permanent link